3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
71
3
Repairs
3.1 Overview
The repair engineer can repair the analyser using standard tools. See the following section for
repairing procedures (including the tools needed). If any repairing step requires a validating step, the
repair engineer should strictly follow the procedure and take the validating step.
When following the repairing procedure as specified in this section, the operator should put on
rubber gloves and clean both hands with disinfectant after performing any repair work.
3.2 Preparatory Work before Repairs
3.2.1 Disassembling the Left Side Panel
Purpose
While performing regular machine maintenance and inspection, the left-side door needs to be
opened to disassemble the sampling syringe assembly, reagent testing panel, or electromagnetic
valve assembly, which are all in the left section of the machine.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Opening
1. As shown in Figure 3-1, use #2 (Ph2) cross screw driver to take out four M3×8 cross recessed
pan-head combination screws on the back panel, and take off the back panel.