EX1629 Multi-Instrument Operation
90-264 VAC
50 / 60 Hz
200VA MAX.
WIDEBAND OUT CH 0-15
WIDEBAND OUT CH 16-31
WIDEBAND OUT CH 32-47
A
C
I
N
P
U
T
TRIGGER
BUS
RESET
ETHERNET
10/100 BASE-T
DIGITAL I/O
VXI Technology
LO
HI
CALIBRATION SENSE
90-264 VAC
50 / 60 Hz
200VA MAX.
WIDEBAND OUT CH 0-15
WIDEBAND OUT CH 16-31
WIDEBAND OUT CH 32-47
A
C
I
N
P
U
T
TRIGGER
BUS
RESET
ETHERNET
10/100 BASE-T
DIGITAL I/O
VXI Technology
LO
HI
CALIBRATION SENSE
90-264 VAC
50 / 60 Hz
200VA MAX.
WIDEBAND OUT CH 0-15
WIDEBAND OUT CH 16-31
WIDEBAND OUT CH 32-47
A
C
I
N
P
U
T
TRIGGER
BUS
RESET
ETHERNET
10/100 BASE-T
DIGITAL I/O
VXI Technology
LO
HI
CALIBRATION SENSE
90-264 VAC
50 / 60 Hz
200VA MAX.
WIDEBAND OUT CH 0-15
WIDEBAND OUT CH 16-31
WIDEBAND OUT CH 32-47
A
C
I
N
P
U
T
TRIGGER
BUS
RESET
ETHERNET
10/100 BASE-T
DIGITAL I/O
VXI Technology
LO
HI
CALIBRATION SENSE
Master
Slave 2
Slave 1
Slave
n
Terminator
Terminator
F
IGURE
A-1:
D
AISY
-
CHAIN
C
ONFIGURATION
NOTE
Due to signal integrity concerns, the maximum number of EX1629s that should be connected in a
daisy-chain configuration is 20.
The Star topology connects the Master and Slave(s) in parallel. In the Star topology, an LXI
Trigger Bus Hub or Switch (e.g., EX2108) is used to distribute the LXI Trigger Bus signals. The
Star topology has two potential advantages over the Daisy-Chain configuration: it can provide
better cable-length matching (and thus signal propagation time matching), and more instruments
can be connected together. In the Daisy-Chain topology, slave devices are connected to the master
at various points along the shared bus, with differing amounts of cable between them and the
master device. This causes signals to reach each of the slaves at slightly different times (a good
rule of thumb is that electrical signals travel along cables at approximately 1 ft/ns).