www.vtiinstruments.com
EX1629 Preparation for Use
39
Set the EX1629 to Static IP
1)
Disconnect all network connections from the PC except for the connection between the PC
and the EX1629.
2)
Apply power to the EX1629 and wait for the LAN LED to turn green.
3)
Discover EX1629 using the LInC-U utility as described in the
in
. The steps taken previously should ensure that discovery works.
4)
Once connected to the embedded web interface, click on the
Network Configuration
link in
the command menu.
5)
Select the
Static
checkbox in the
IP Address Source
field and ensure that
DHCP
and
AutoIP
are not selected.
6)
In the
IP Address
field, enter an appropriate IP address for the EX1629. Use the IP address
obtained the
Determine PCs Network Settings
step to determine the network address. The
network address is the first three digits of the IP address (10.1.0 in the example provided).
The last digit of the IP address (the node), is a number, 0 through 255, that is not currently
assigned to any other device on the network. Ensure that a unique IP address is assigned to the
EX1629 by consulting a network administrator.
7)
In the
Subnet Mask
field, enter the subnet mask in the
Determine PCs Network Settings
step.
8)
In the
Gateway Address
field, enter the same gateway address in the
9)
Click
Submit
to save all changes made to the
Network Configuration
page.
F
IGURE
2-13:
EX1629
S
ET TO
U
SE
S
TATIC
IP
Restore the Host PCs Network Settings
1)
Power down the EX1629 and reconnect it to the LAN in its desired location.
2)
If the PC was originally set to use a static IP address, use the procedure in
to change the PCs IP address back to its original state.
3)
Power on the EX1629 and wait for the LAN LED to turn solid green.
4)
Discover the EX1629 using VTI’s LInC-U as described in
Using Multiple Network Cards
When multiple network cards exist in a single PC, it may be necessary to define a static IP address
to both the host PC NIC card that will interface with the EX1629 mainframe as well as the
EX1629 itself. This process is only necessary if a DCHP server is not connected to the network to
which the device is connected and typically occurs when the NIC is connected directly to the