EX1629 Command Set
255
vtex1629_set_euconv_excitation
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_set_euconv_excitation (ViSession
vi
, ViInt32 _VI_FAR
channels[]
,ViInt32
numberOfChannels
, ViReal64
euConversionVoltage
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
channels
= an input integer array containing a list of channel numbers for which the EU conversion excitation
voltage will be set. Valid input values: 0 to 47.
numberOfChannels
= the size of the
channels
list. Valid input values: 1 to 48.
euConversionVoltage
= a real input value, in volts, that indicates the value that should be used in EU conversions
for the excitation voltage. Valid input values: 0.00000 to +16.00000.
D
ATA
I
TEM
R
ESET
V
ALUE
euConversionVoltage
= 0.000000
D
ESCRIPTION
This function sets the excitation voltage to be used in EU conversions for a particular list of channels.
NOTE
The conventional method of providing a non-nominal value of the excitation voltage to the EU
conversion
is
to
conduct
an
excitation
voltage
measurement
using
the
vtex1629_measure_excitation_voltage function. This function provides a manual method that is
normally only used for system diagnostic purposes.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
ViInt32 channels[] = {0};
ViInt32 numberOfChannels = 1;
…
status = vtex1629_set_euconv_excitation(instrumentHandle,
channels,
numberOfChannels,
2.0);