www.vtiinstruments.com
EX1629 Command Set
221
vtex1629_self_cal_get_status
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_self_cal_get_status (ViSession
vi
, ViPInt32
percentComplete
, ViPInt32
calStatus
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
percentComplete
= an integer return value, from 0 to 100, indicating a percentage of completion for the self-
calibration process.
calStatus
= an integer return value indicating the status of the currently running self-calibration process, or the last
self-calibration process that was run, and completed. Expected data return values: 0, 1, or 2.
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function returns the status of the self-calibration process. This will be the status of the currently running self-
calibration routine, or the last self-calibration routine that was completed.
The
calStatus
parameter indicates the following:
0 = Self-calibration in progress
1 = Self-calibration complete
2 = Self-calibration failed.
NOTE
Additional instrument driver calls should not be performed until the result of
calStatus
is equal to
VTEX1629_SELF_CAL_COMPLETE.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
ViInt32 percent;
ViInt32 calstatus;
…
status = vtex1629_self_cal_get_status (instrumentHandle, &percent, &calstatus);