VTI Instruments Corp.
84
EX1629 Programming
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_ONE,
VTEX1629_LXI_INPUT,
VTEX1629_LXI_INTERNAL_EXTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_TWO,
VTEX1629_LXI_INPUT,
VTEX1629_LXI_INTERNAL_EXTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_THREE,
VTEX1629_LXI_INPUT,
VTEX1629_LXI_INTERNAL_EXTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_FOUR,
VTEX1629_LXI_OUTPUT,
VTEX1629_LXI_INTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_FIVE,
VTEX1629_LXI_OUTPUT,
VTEX1629_LXI_INTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_SIX,
VTEX1629_LXI_OUTPUT,
VTEX1629_LXI_INTERNAL );
result = vtex1629_set_lxibus_configuration( slave_vi,
VTEX1629_LXI_LINE_SEVEN,
VTEX1629_LXI_OUTPUT,
VTEX1629_LXI_INTERNAL );
/* Configure the sample clock on the slave as an input from LXI0. */
result = vtex1629_set_sample_clock_source( slave_vi,
VTEX1629_SAMP_CLK_MODE_SLAVE,
VTEX1629_LXI_LINE_ZERO,
VTEX1629_LXI_LINE_NONE );
/* Configure the synch source on the slave as an input from LXI1. */
result = vtex1629_set_synch_source( slave_vi,
VTEX1629_SYNC_MODE_SLAVE,
VTEX1629_LXI_LINE_ONE,
VTEX1629_LXI_LINE_NONE );
/* Configure the trigger source on the slave as an input from LXI2.
*/
result = vtex1629_set_trigger_source( slave_vi,
VTEX1629_TRIG_SRC_LXI2_POS );
/* Configure the ARM source on the slave as an input from LXI3. */
result = vtex1629_set_arm_source( slave_vi,
VTEX1629_TRIG_SRC_LXI3_POS );
/* Issue a soft synch command to the master to generate a
synchronization signal since the sample clock source was changed. */
result = vtex1629_soft_synch( master_vi )
Содержание EX1629
Страница 310: ...VTI Instruments Corp 310 EX1629 Command Set...
Страница 342: ...VTI Instruments Corp 342 EX1629 Onboard Memory...