VTI Instruments Corp.
32
EX1629 Preparation for Use
TEDS
COMM
External customer
shunt cal
resistor terminals
Internal
Rcal
55K
-R Cal
+R Cal
Pin 5
Pin 4
Strain
Bridge
8-pin telcom
connector (RJ-45)
350
120
User
10K
10K
+Excitation Sense
+ Excitation
- Excitation
Full-Bridge
(1 of 48 channels)
+Excitation
+Sense
-Sense
+Excitation Sense
-Excitation Sense
-Excitation
-Excitation Sense
RJ-45 Pin#
12345678
Pin 2
Pin 7
Pin 6
Pin 3
Pin 8
Pin 1
+
-
-
+
10K
10K
Instrumentation
Amplifier
+
-
x1
x10
x100
Hi
Lo
Shield
150
150
-Sense
-Cal
AGND
+Sense
Wagner
+Cal
AGND
430 kHz
LOW-PASS
FILTER
60.3 kHz
LOW-PASS
FILTER
ADC
TRIFILAR
FILTER
F
IGURE
2-5:
F
ULL
-B
RIDGE
C
ONFIGURATION
Furthermore, it is critical to understand how the EX1629 defines its measurement in this
configuration. Specifically, it defines the measured strain to be the strain that is present in each
gage, not the total of the four gages. For example, if the positive gages are subjected to tensile
strain of 1000 µε and the negative gages are subjected to compressive strain of 1000 µε, the
EX1629 will measure a value of +1000 µε, not +4000 µε.
V
OLTAGE
M
EASUREMENT
C
ONFIGURATIONS
The EX1629 main input channels can also be used for general voltage measurement. For this
application, the channel is effectively configured for Full-Bridge measurements (i.e., no
completion resistor or “back-half” of the bridge is enabled). The signal to be measured should be
connected to the +Sense and –Sense lines. The excitation source may be used in this mode, if the
measurement requires it, but the excitation source limits should be kept in mind.
For fully differential inputs, the configuration in Figure 2-6 should be used. For situations that
require a single-ended connection (i.e. one side grounded), the configuration in Figure 2-7 should
be used. When using these configurations, Negative Excitation Sense should be set to 0 V.