www.vtiinstruments.com
EX1629 Command Set
307
Error Code Error Message
Meaning
FFBFFC0A91 VTEX1629_ERROR_DIO_CONFLICT_DUPLICATE_ENTRY
Requested event configuration conflicts
with existing configuration for the input
event and/or output action specified.
FFBFFC0A92 VTEX1629_ERROR_DIO_INVALID_INPUT_LINE
Invalid digital input line.
FFBFFC0A93 VTEX1629_ERROR_DIO_INVALID_OUTPUT_LINE
Invalid digital output line.
FFBFFC0A94 VTEX1629_ERROR_DIO_INVALID_INPUT_EVENT
Invalid input event.
FFBFFC0A95 VTEX1629_ERROR_DIO_INVALID_OUTPUT_ACTION
Invalid output action.
FFBFFC0A96 VTEX1629_ERROR_QTR_BRIDGE_NOT_SELECTED
Channel is not in quarter bridge
configuration.
FFBFFC0A97 VTEX1629_ERROR_TRIGINIT_NEED_SOFTSYNC
Triginit needs softsync first.
FFBFFC0A98 VTEX1629_ERROR_FS_NOT_SUPPORTED_FOR_DYNAMIC_EXCITATION_EU
Sampling frequency greater than 1 kHz
not supported while in current mode.
FFBFFC0A99 VTEX1629_ERROR_DYNAMIC_EXCITATION_EU_NOT_SUPPORTED_AT_CURR_FS
This mode is not supported while
sampling rate is greater than 1 kHz.
FFBFFC0A9A VTEX1629_ERROR_DYNAMIC_EXCITATION_EU_CONF_EXCITEOUT_SRC_NOT_ENABLED The confidence sources
VTEX1629_CONFSRC_EXCITE_POS
and
VTEX1629_CONFSRC_EXCITE_NEG
must be in the confidence scanlist in
order to enable dynamic excitation EU.
FFBFFC0A9B VTEX1629_ERROR_DYNAMIC_EXCITATION_EU_NEEDS_CONF_EXCITEOUT_SRC
Can not disable confidence excitation
source while dynamic EU conversion is
in progress.
FFBFFC0A9C VTEX1629_ERROR_CONF_LIMIT_INVALID_REPORTING_TYPE
Invalid reporting type.
FFBFFC0A9D VTEX1629_ERROR_BRIDGE_LIMIT_NOT_SUPPORTED_AT_CURR_FS
Can not enable bridge limit checking
when the bridge sampling frequency is
greater than 1 kHz.
FFBFFC0A9E VTEX1629_ERROR_FS_NOT_SUPPORTED_FOR_BRIDGE_LIMIT_CHECKING
Can not set sampling frequency greater
than 1 kHz while in the current mode.
FFBFFC0A9F VTEX1629_ERROR_CONF_LIMIT_NOT_SUPPORTED_AT_CURR_FS
Summary reporting or detailed reporting
not supported for sampling frequency
greater than 1 kHz.
FFBFFC0AA0 VTEX1629_ERROR_FS_NOT_SUPPORTED_FOR_CONF_LIMIT_CHECKING
Can not set sampling frequency greater
than 1 kHz while in current mode.
FFBFFC0AA1 VTEX1629_ERROR_HW_DOES_NOT_SUPPORT_CONF_EXCITEOUT_BUFF_SRC
Current hardware prohibits enabling
EXCITEOUT_BUFF. This is only
available on newer hardware revisions.
FFBFFC0AA2 VTEX1629_ERROR_CAL_FILE_NOT_PRESENT
Calibration file not present.
FFBFFC0AA3 VTEX1629_ERROR_CAL_FILE_TOO_LARGE
Calibration file too large.
FFBFFC0AA4 VTEX1629_VXI11_ERROR_NO_ERROR
VXI11 No error.
FFBFFC0AA5 VTEX1629_VXI11_ERROR_SYNTAX_ERROR
VXI11 Syntax error.
FFBFFC0AA6 VTEX1629_VXI11_ERROR_DEVICE_NOT_ACCESSIBLE
VXI11 Device not accessible.
FFBFFC0AA7 VTEX1629_VXI11_ERROR_INVALID_LINK_IDENTIFIER
VXI11 Invalid link identifier.
FFBFFC0AA8 VTEX1629_VXI11_ERROR_PARAMETER_ERROR
VXI11 Parameter error.
FFBFFC0AA9 VTEX1629_VXI11_ERROR_CHANNEL_NOT_ESTABLISHED
VXI11 Channel not established.
FFBFFC0AAA VTEX1629_VXI11_ERROR_OPERATION_NOT_SUPPORTED
VXI11 Operation not supported.
FFBFFC0AAB VTEX1629_VXI11_ERROR_OUT_OF_RESOURCES
VXI11 Out of resources.
FFBFFC0AAC VTEX1629_VXI11_ERROR_DEVICE_LOCKED_BY_ANOTHER_LINK
VXI11 Device locked by another link.
FFBFFC0AAD VTEX1629_VXI11_ERROR_NO_LOCK_HELD_BY_THIS_LINK
VXI11 No lock held by this link.
FFBFFC0AAE VTEX1629_VXI11_ERROR_IO_TIMEOUT
VXI11 IO timeout.
FFBFFC0AAF VTEX1629_VXI11_ERROR_IO_ERROR
VXI11 IO error.
FFBFFC0AB0 VTEX1629_VXI11_ERROR_INVALID_ADDRESS
VXI11 Invalid address.
FFBFFC0AB1 VTEX1629_VXI11_ERROR_ABORT
VXI11 Abort.
FFBFFC0AB2 VTEX1629_VXI11_ERROR_CHANNEL_ALREADY_ESTABLISHED
VXI11 Channel already established.
FFBFFC0AB3 VTEX1629_ERROR_BRIDGE_LIMIT_CHECKING_NOT_SUPPORTED_AT_CURR_FS
Bridge limit checking not supported at
current sampling frequency.
FFBFFC0AB4 VTEX1629_ERROR_FS_NOT_SUPPORTED_WHEN_BRIDGE_LIMIT_CHECKING_ENABLED Sample frequency not supported when
bridge limit checking is enabled.