VTI Instruments Corp.
276
EX1629 Command Set
vtex1629_set_sample_clock_source
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_set_sample_clock_source (ViSession
vi
, ViInt32
sampleClockMode
, ViInt32
inLine
, ViInt32
outLine
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
sampleClockMode
= an integer input value that determines whether the EX1629 is operating as a master or slave.
Valid input values: VTEX1629_SAMP_CLK_MODE_MASTER or VTEX1629_SAMP_CLK_MODE_SLAVE.
inLine
= an integer input value that determines the trigger bus line configured to use for sample clock input. Valid
input values: VTEX1629_LXI_LINE_ZERO, VTEX1629_LXI_LINE_FOUR, or VTEX1629_LXI_LINE_NONE.
outLine
= an integer input value that determines the trigger bus line configured to output sample clock events. Valid
input values: VTEX1629_LXI_LINE_ZERO to VTEX1629_LXI_LINE_SEVEN, VTEX1629_LXI_LINE_NONE.
D
ATA
I
TEM
R
ESET
V
ALUE
inLine
= VTEX1629_LXI_LINE_NONE
outLine
= VTEX1629_LXI_LINE_NONE
sampleClockMode
= VTEX1629_SAMP_CLK_MODE_MASTER
D
ESCRIPTION
This function sets the sample clock source for the device.
The
sampleClockMode
parameter indicates whether the EX1629 is configured as a master device that outputs a
sample clock for itself and other devices or as a slave device that receives its sample clock from another device.
When operating in standalone mode,
sampleClockMode
should be configured as a master.
The
inLine
parameter indicates the LXI line that should be used as the sample clock input. This value is applicable
regardless of whether the device is configured as a master or a slave. When
inLine
is set to
VTEX1629_LXI_LINE_NONE, a special, internal sample clock line is used.
The
outLine
parameter indicates the LXI line that should be used as the sample clock output. This value is only
applicable when the device is configured as a master. When
outLine
is set to VTEX1629_LXI_LINE_NONE, the
sample clock is output on a special, internal sample clock line. When in Slave mode, no output line can be used, so
VTEX1629_LXI_LINE_NONE must be specified.
Decimal
Value
Hex
Value
#define Symbol
inLine/outLine
Description
0
0x00
VTEX1629_LXI_LINE_ZERO
LXI LINE 0
4
0x04
VTEX1629_LXI_LINE_FOUR
LXI LINE 4
8
0x08
VTEX1629_LXI_LINE_NONE
None
When in master mode, the
inLine
and
outLine
values may be the same or they may be different. One case where
they would be different is if the master is outputting the sample clock on one LXI line and receiving it back in from
an LXI Trigger Bus hub on another line. When in standalone mode,
inLine
and
outLine
will always be the same.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
…
status = vtex1629_set_sample_clock_source(instrumentHandle,
VTEX1629_SAMP_CLK_MODE_MASTER,
VTEX1629_LXI_LINE_ZERO,
VTEX1629_LXI_LINE_ZERO);