Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
133 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
8) Click the zoom icon and drag right to magnify the line pair of Quart Phantom. Adjust
BRIGHTNESS and CONTRAST bar to show the line pair well.
9) [Line Pair Resolution] Make sure that the lines can be seen clearly at 2.5 lp/mm.
10) [Low Contrast Resolution] Make sure that a 2.0mm hole can be seen clearly.
If the standard is not met,
- Refer to the PANO Geometry Calibration section and perform CENTER BALL Size
(PYST) Calibration.
- Perform PANO sensor calibration and PANO Collimator Alignment again.
Содержание Smart Plus PHT-35LHS
Страница 1: ...Technical manual Model PHT 35LHS Version 1 16 English...
Страница 2: ......
Страница 158: ...Smart Plus PHT 35LHS Technical Manual VT TECH Smart Plus 001 156 188 COPYRIGHT 2018 VATECH ALL RIGHTS RESERVED...
Страница 190: ...www vatech com...