Ex) [SPM_FANT_0400] = When the Tube Tank temperature is over
40
℃
, FAN is turned on.
[SPM_PPON_000#]
Set to enable/disable Power Profile (Only PANO mode)
(# = 0: Disable Power Profile): Default
(# = 1: Enable Power Profile)
: With this command, Power Profile is automatically enabled after
restarting the system.
Canine Laser Beam Control Command
Command
Description
[SPM_CHST_####]
Set Reference position value of Canine Laser Beam
(Initial-Value: 0380 (1~999))
[SPM_CHV?]
Request Current reference position value of Canine Laser Beam
[SPM_CENA_000#]
Set to enable/disable Canine Laser Beam
(# = 0: Disable Canine Laser Beam)
(# = 1: Enable Canine Laser Beam)
X-Ray Exposure Count Command
Command
Description
[SPM_GCNT]
Request X-Ray exposure count (All capture mode)
[SPM_PCNT]
Request X-Ray exposure count (PANO capture mode)
[SPM_ACNT]
Request X-Ray exposure count (ARCT capture mode)
[SPM_HCNT]
Request X-Ray exposure count (CEPH capture mode)
[SPM_ICNT_000#]
Initialize X-Ray exposure count
(# = 0: Initialize All capture count)
(# = 1: Initialize PANO capture count)
(# = 2: Initialize CBCT capture count)
(# = 3: Initialize CEPH capture count)
(# = 4: Initialize ARCT capture count)
Содержание Smart Plus PHT-35LHS