78 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
9) Type the mA value that meets the standard of the Background Level in the High-Resolution
section. For Woman, subtract 0.3mA from the mA in Man mode and for Child, subtract 0.6mA
from the mA in Man mode and type the value in the section.
•
Do not change the kVps.
•
Make sure that each FOV (FOV 0, FOV 1, FOV 2) have the same values
By MAN_HARD=0950, 0870, ‘0950’ means 95kVp and ‘0870’ means 8.7mA.
10) Save and exit PwrParam_CBCT.ini.
ex) 8.7mA
ex) 8.4mA = 8.7mA - 0.3mA
ex) 8.1mA = 8.7mA - 0.6mA