Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
132 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
4) Adjust Canine Laser Beam knob to align the Beam to the baseline of Quart Phantom as shown
below.
Canine Laser Beam must be aligned with the center of the Quart Phantom AL filter.
as shown in the figure
5) Attach 1.8 mm (or 1.0mm + 0.8mm) Copper Filter on X-Ray Tube as shown below.
6) Click READY and acquire and save the image according to the Console Software guidance.
7) Open the saved image in the 2D Viewer.