Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
64 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
12) If the red pixel count is out of the permitted range, change the top/bottom collimator position
values.
0.1(mm) is approximately 3 pixels.
To close the 4-axis collimator, decrease the collimator position values
To open the 4-axis collimator, increase the collimator position values.
13) Do the steps 3) through 10) until the top/bottom red pixel count is within the permitted value.
14) When the Collimator Alignment at 100 x 85 is completed, click the “CALC” button. FOV 100 x
70, 50x50 values will be calculated.
15) When the Collimator Alignment is completed, click Close to exit the Collimator Align program.
○
1
Содержание Smart Plus PHT-35LHS
Страница 1: ...Technical manual Model PHT 35LHS Version 1 16 English...
Страница 2: ......
Страница 158: ...Smart Plus PHT 35LHS Technical Manual VT TECH Smart Plus 001 156 188 COPYRIGHT 2018 VATECH ALL RIGHTS RESERVED...
Страница 190: ...www vatech com...