Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
86 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
15) When CT Number Calibration is completed, open Reconstruction. xv.
(VAKCAP.exe -> Settings tab -> Double-click Reconstruction.vxm).
16) Recover the parameters as shown below and save/exit the file.
•
POSTPROCESSING = 0 → POSTPROCESSING =
-50
•
UserCTNumCal = 0 → UserCTNumCal = 1
•
UseArcCut = 0 → UseArcCut = 1