Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
181 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
CT Align & Calibration Setting Command
Command
Description
[SPM_AXST_0000]
Set X axis Reference position of CT mode
Initial-Value: 0103(1~1999)
#### = dyyy (d: Direction, yyy: Distance) ex. 0234 = 2.34cm
[SPM_AX?_]
Request current “AXST” Parameter value
[SPM_AYST_0000]
Set X axis Reference position of CT mode
Initial-Value: 0160(1~1999)
#### = dyyy (d: Direction, yyy: Distance) ex. 0234 = 2.34cm
[SPM_AY?_]
Request current “AYST” Parameter value
[SPM_COFS_#####]
Set Capture starting position of Rotating Unit in CT mode
(##### = If the first digit is 1, it means clockwise, and 0 means
counter-clockwise)
Initial-Value: 00001 (1~19999)
[SPM_CO?_]
Request current “COFS” Parameter value
[SPM_ACHF_####]
Set CT Alignment Half Reference Value
Initial-Value: 1200 (1~9999)
Check Command
Command
Description
[SPM_ECHK]
Check Exposure Switch
If the exposure switch is pressed, it replies [epm_echk_002]
If not, it replies [epm_echk_001]
[SPM_DCHK]
Check the state of the X-Ray Room Door in Console Software
(If Door is closed, it replies [epm_dchk_001])
(If Door is opened, it replies [epm_dchk_002])
※
If the Door Lock Option is “Disable,” it replies [epm_dchk_001]
Etc. Command
Command
Description
[SPM_RINT]
Initialize Parameter value of MCU Board
Caution: This command initializes all Parameter values saved in
MCU
Содержание Smart Plus PHT-35LHS
Страница 1: ...Technical manual Model PHT 35LHS Version 1 16 English...
Страница 2: ......
Страница 158: ...Smart Plus PHT 35LHS Technical Manual VT TECH Smart Plus 001 156 188 COPYRIGHT 2018 VATECH ALL RIGHTS RESERVED...
Страница 190: ...www vatech com...