If the median value does not meet the requirement, do the following steps:
①
Click the Settings tab.
②
Adjust kVp0 value by 10 in the [BrightCalibration] field.
③
Click Save.
④
Input the following commands only if you restarted the equipment, otherwise continue to
step V below: [SPM_ARCT] [SPM_CSON] [SPM_FISS_0001] [SPM_FIPM_0001]
[SPM_FISM_0001] [SPM_FOVO] [SPM_COLM_0009]
⑤
Acquire Bright calibration data again.
21) Click the 2 Point Bright Calibration data file on View16 and ensure that
the Median value meets the standard. If the median value does not meet the requirement,
do the following steps:
①
Click the Settings tab.
②
Adjust kVp1 value by 10 in the [BrightCalibration] field.
③
Click Save.
④
Input the following commands only if you restarted the equipment, otherwise continue to
step V below [SPM_FISS_0001] [SPM_FIPM_0001] [SPM_FISM_0001] [SPM_FOVO]
[SPM_COLM_0009] [SPM_ARCT] [SPM_CSON]
⑤
Acquire Bright calibration data again.
22) Copy and paste the 2-Point Bright Calibration data to the following folder:
C:\VCaptureSW\Acquisition\CBCT\WidePano\CAL\Open
23) Attach Copper (1mm) Filter to obtain Bright Calibration data of 4Point.
2. Check kVp, mA value.
4. When kVp, mA value
is checked, click
Save
1. Check the NumPoints=4
3. Check comment processing.