7) When the Raw Viewer window appears, release the X-ray exposure switch.
8) Select 0009. raw in the raw file list.
9) Type 70 in the T box.
10) Select S in the M box.
11) Make sure that the result meets the PANO collimator-alignment standard.
Position
Permitted value
(Red pixels covered by collimator)
Top
10 ~ 35 pixels
Bottom
10 ~ 25 pixels
Left
2 ~ 4 pixels
Right
2 ~ 4 pixels
Top
Bottom
Left
Right