1.3. Performing CEPH Sensor Bright Calibration
Follow the procedure below to perform CEPH sensor bright calibration.
1) Sensor Bright Calibration is acquired in Norm (same as Fast) mode.
2) Attach Cu (1mm) Filter to obtain 5 points bright Cal.
3) Run the C:\VCaptureSW\Acquisition\CEPH \WideCEPH \Norm(Fast)\VACAP.exe file.
4) The standard of the level value for each bright calibration point is as follows:
※
Bright standard value: 5 points with Cu 1mm filter
Bright
Min
Max
Median Value
Cal Point 1
50
80
Cal Point 5
400~