Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
41 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
6) Adjust the tripod to constant height and put the tripod and the BEAD PHANTOM in position as
shown in the figure.
7) Run VAKCAP.exe. and input the following commands:
[SPM_ARCT] [SPM_CSON]
[SPM_FISS_0001] [SPM_FIPM_0001] [SPM_FISM_0001]
8) Select
Capture/Recon
tab and type
[SPM_FOVO]
click Send to initiate the system into CT
mode, then type
[SPM_COLM_0009]
to provide the power.
9) Type
[SPM_AAMS]
and click Send to go into CT capture mode A (at 15 secs, 360 degrees).
10) Select the
Soft Calib
tab and click
Dark
to acquire Dark Calibration Data.
11) Click the
Capture/Reco
n tab and set the parameters as shown below.
12) Click
Capture
and make the exposure by pressing the X-ray exposure switch button.
Leave the radiology room to minimize radiation exposure when the X-ray system starts
operation after you press the button.
2~5mm