13) Click CBCT tab, type average value in each box for the FOV1 under HIGH RESOLUTION.
Click
Save
to keep the changes, then click
Close
.
Do not measure the Metal value and use the fixed value of 2300 for every FOV.
14) Do steps 2) through 14) for other modes as follows:
Count
FOV
Image Quality
Voxel
Gender
Tooth type
100 x 85(FOV.1)
Green
Standard (0.2)
Man
Hard
100 x 70(FOV.2)
High Resolution
Standard (0.2)
Man
Hard
100 x 70(FOV.2)
Green
Standard (0.2)
Man
Hard