11) After completing the CEPH Calibration, select the
①
CEPH Processing button, as shown in
[8].
12)
②
:When the Mode Confirm button is selected, Level 1~3 images are created.
* Level 1~3: Level 3 has the highest contrast due to the difference in contrast.
13) Select the O mark on the far right of
③
, which is an image of Level 3 among the created
images.
14)
④
When the Simulation button is selected, an image of O marked in red, as shown in [9], is
created.
15)
⑤
Select O marked in red and
⑥
click the Apply button.
16) When a pop-up window is created, as shown in [10],
⑦
click the OK button.
17) As shown in [11], check that the time of the modified date of Acquisition / WideCEPH / Norm
or Fast / UseParam / ZeusCEPH is the same as the time when the work is completed.