Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
117 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
10) Click V to open View16
11) Click the 1 Point Bright Calibration data file on View16 and make sure that the Median value
meets the standard.
If the median value does not meet the requirement, do the steps that follow:
①
Click the Settings tab.
②
Adjust kVp0 value by 10 in the [BrightCalibration] field.
③
Click Save.
④
Input the following commands only if you restarted the equipment, otherwise continue
to step V below: [SPM_PANO] [SPM_CENA_0001] [SPM_FISS_0001]
[SPM_FIPM_0004] [SPM_FISM_0001] [SPM_COLM_0008] [SPM_FOVO]
[SPM_PSON]
⑤
Acquire Bright calibration data again.
12) Click the 5 Point Bright Calibration data file on View16 and make sure that the Median value
meets the standard.
If the median value does not meet the requirement, do the steps that follow:
①
Click the Settings tab.
②
Adjust kVp1 value by 10 in the [BrightCalibration] field.
③
Click Save.
④
Input the following commands only if you restarted the equipment, otherwise continue
to step V below [SPM_PANO] [SPM_CENA_0001] [SPM_FISS_0001]
[SPM_FIPM_0004] [SPM_FISM_0001] [SPM_COLM_0008] [SPM_FOVO]
[SPM_PSON]
⑤
Acquire Bright calibration data again.
13) When the calibration is complete, remove the 1 mm Copper Filter.
max
ave
med
min
132
66.4
66
0