5.CALIBRATION
44
5-7-11. Speed input count
When “NORMAL MODE A”, “NORMAL MODE B”, “NORMAL A SLOW”, “ANA.
NORMAL A”, or “ANA. A SLOW” is selected in [SPEED ENTRY SELECT], specify
a pulse count (9-digit integer) that represents the travel length of the belt (BELT
LENGTH).
For example, if the [BELT LENGTH] is set to 10m, and the rotary encoder generates
25 pulses during a travel of 10m, enter “25” in this field.
When PHASE MEASURE is selected in [SPEED ENTRY SELECT], specify a count
of rotation (9-digit integer, decimal place is fixed to 2
nd
digit) that corresponds to the
travel length (BELT LENGTH).
For example, if the [BELT LENGTH] is set to 10m, and the rotary encoder rotates 3.70
times per 10m, enter “3.70” in this field.
Operation
5-7-12. Error meas.select
You can select a reference value used to stop the zero or span error measurements
(ZERO ERROR MEASURE/SPAN ERROR MEASURE).
If you select “BELT LENGTH”, the measurement stops when the travel length of the
belt, from the start of error measurement, exceeds the value specified by [BELT
LENGTH].
If you select “SPEED INPUT COUNT”, the measurement stops when the speed pulse
count from the start of error measurement exceeds the value specified by [SPEED
INPUT COUNT].
Operation
Range: 1 – 999999999 (9 digits)
0.01 – 9999999.99 (9 digits, decimal place fixed)
MODE
FLOW CALIBRATION
→
→
SPPED INPUT COUNT
PAGE
→
Options: TESTING BELT, SPEED COUNT
MODE
FLOW CALIBRATION
→
→
ERROR MEAS.SELECT
PAGE
→