12
Name of parts and functions
3.2
●
Rear panel
⑰
Protecting grounding
Terminal for grounding to the earth.
terminal
Make sure to ground to the earth using the attached earth cable (green).
It is the same voltage as the case of the tester.
⑱
FUSE 10A
Fuse socket. The rate of fuse is as the following table shows.
Type
Power source voltage
Rate of fuse
Standard
100V AC
125V 10A
Optional
115V AC
200V AC
250V 5A
220V AC
240V AC
Please do not use it excluding an attached fuse because it is a rush-resistive
type.
⑲
100V~50/60Hz
Inlet for connection of supply power source.
It conforms to the attached power cord (3P).
⑳
RS-232C
Connector for RS-232C serial communication (D-sub 9 pins).
Refer to the instruction manual of interface.
REMOTE/OUT
Connector for the setting inputs of interlock and to output the output signals.
For detail, refer to the article 15.1 (P52).
HIGH VOLTAGE
High voltage side terminal of test voltage output.
It outputs high voltage during the test, so never touch it during the
DANGER
lamp
⑬
is lit up. The operator may suffer electric shock.
It is common with
HIGH VOLTAGE
on the front panel.
LOW
Low voltage side terminal of the test voltage output. It is of the same
voltage as the case of this tester.
VOLTAGE MONITOR
Monitor output of withstanding voltage output.
Output voltage: 0~5V DC (to 0~5kV AC)
100V
50/60Hz
C
C
C
REAR:ON
STOP
START
REMOTE
a
c
NC
STATUS OUT
注
意
CAUT ION
PROTECT IVE EARTH TERMINAL .
ENSUREGROUNDINGOF
大地アースして ください 。
保護接地端子は確実に
RS-232C
REMOTE/OUT
LOW
HIGH VOLTAGE
+
-
VOLTAGE MONITOR
OUTPUT
MAX 5kV
17
19
20
21
22
23
24
25
26
FUSE
T10AH 125V
18