24
Setting of test condition for withstanding voltage test
[When setting the referential voltage]
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks.
Test mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
W-I
,
I-W
or
W
lamp).
②
Press
or
key and make the test mode lamp lit.
Setting of referential voltage
①
Press
or
key and select the status that the
W AC
lamp is lit up, the test
voltage range display is lit up with
and the test voltage display blinks.
②
Next, press
ON/OFF
key and select the status that the display blinks with the
numeral.
③
Press
▲
or
▼
key and set the referential voltage.
Pressing of
key (
SHIFT
and
▲
at a time) or
key (
SHIFT
and
▼
at a time) allows the setting of second digit (the digit of 0.10kV)
(refer to the above figure).
Note
: The referential voltage can be set within the range of 0.00~5.00kV.
To move to the previous setting
Press
key, then the setting changes to the
setting of slider voltage
.
To the next setting
Press
key, then changes to the
setting of high limit of leak current
.
Finish of setting
Press
ENTER
key, then the slider moves up or down and the tester returns to READY
status, memorizing the set having been made.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the setting mode for the test condition is interrupted and the tester becomes READY
status.
The test conditions in this case are those before entering the setting mode of test condition.
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
M
Ω
LOW
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
○
○
○
I
I-TEST
I
I-W
W-I
mA
HIGH
●
●
VOLTAGE SLIDER