45
I-W test
11.4
●
I-W test (insulation resistance
→
withstanding voltage test)
In READY status, the display of test conditions of withstanding voltage test and insulation
resistance test alternates at the cycle of 2 seconds.
(1) In READY status, press
or
key, then the test mode lamp blinks.
Select the
I-W
lamp with
▲
or
▼
key, and the test conditions are also displayed.
(2) When it is necessary, the test conditions can be changed with
,
,
▲
,
▼
,
ON/OFF
,
SHIFT
key. For detail, refer to the article 8 and 9.
(3) Press
START
switch
③
, then
I-TEST
lamp is lit up and the test starts.
While the high voltage is output, the
DANGER
lamp is lit up.
During the test,
TEST/H.V.OUT
,
TEST
of
REMOTE/OUT
connector
are turned ON
and
READY
is OFF.
(During the insulation resistance test,
I-TEST
is, and during the withstanding voltage
test,
W-TEST
is turned ON.)
When the test is finished,
TEST/H.V.OUT
,
W- TEST
,
I-TEST
,
TEST
of
REMOTE/OUT
connector
are turned OFF and
END
is ON. For detail, refer to the article 17
Timing chart.
(4) Case of good judgement
When the test time of insulation resistance test or withstanding voltage test is elapsed,
GOOD
of insulation resistance test or withstanding voltage test is respectively lit up.
●
When the good judgement time is for 0.2 seconds, the tester returns to READY
status before the start.
●
When the good judgement time is continuous output, re-start is not possible.
In this case, press
STOP
switch
②
, then it becomes READY status.
●
For the status output, refer to the article 17 Timing chart.
(5) Case of NG judgement
When the insulation resistance test is NG
- Refer to the article 11.2
When the withstanding voltage test is NG
- Refer to the article 11.1
Press
STOP
switch
②
, then it becomes READY status.
○
W
READY
REMOTE
DISCHARGE
KEY LOCK
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
HIGH
LOW
GOOD
HIGH
LOW
W-TEST
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
○
○
I
I-TEST
I
I-W
W-I
●
●
○
HIGH
mA
LOW
M
Ω
GOOD
○
○
●
●
○
○
W
READY
REMOTE
DISCHARGE
KEY LOCK
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
M
Ω
LOW
HIGH
LOW
HIGH
LOW
○
W-TEST
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
○
○
○
○
I
I-TEST
I
I-W
W-I
mA
HIGH
GOOD
GOOD
○
●
●