29
Explanation of command
4.2.39
SET:
(setting of parameters of test condition)
Function
Makes the setting of test mode and parameters in the lump.
Structure
SET:
Parameter of test
Parameter of test
MODE=
Test mode
W-I
,
I-W
,
W
,
I
WMARK=
Setting is made for the test mode which includes
WLEVEL=
the withstanding voltage test.
W-I
,
I-W
,
W
WHIGH=
For detail, refer to the articles 4.2.15 (P16),
WLOW=
4.2.17 (P17), 4.2.19 (P18), 4.2.21 (P19) and
WTIMER=
4.2.23
(P20).
IVOLT=
Setting is made for the test mode which includes
IHIGH=
the insulation resistance test.
W-I
,
I-W
,
I
ILOW=
For detail, refer to the articles 4.2.25 (P21),
IMASK=
4.2.27 (P22), 4.2.29 (P23), 4.2.31 (P24),
ITIMER=
4.2.33 (P25) and 4.2.35 (P26).
DISCHARGE=
Note
: In the test mode W or I, omit the items which are not tested.
Transmission
When the test mode is W-I,
SET:MODE=WI, WMARK=2.50kV, WLEVEL=1.50kV, WHIGH=20.0mA, WLOW=OFF,
WTIMER=60.0s, IVOLT=0.5kV, IHIGH=OFF, ILOW=10MOHM, IMASK=1.0s,
ITIMER=60.0s, DISCHARGE=ON
When the test mode is I,
SET:MODE=I, IVOLT=0.5kV, IHIGH=OFF, ILOW=10MOHM, IMASK=1.0s,
ITIMER=60.0s, DISCHARGE=ON
Response
When 8527 received the effective command setting.
ERROR=0
…….…….
When the Response Setting is ON.
No response ……………..
When the Response Setting is OFF.
4.2.40
SET:?
(lump read-out of parameters of test condition)
Function
Reads out the test mode and each parameter in the lump.
Structure
SET:?
Transmission
SET:?
Response
When the test mode is I-W. (
MODE=WI
when the mode is W-I)
○
When
FORMAT=ON
SET:MODE=IW, WMARK=2.50kV, WLEVEL=1.50kV, WHIGH=20.0mA, WLOW=OFF,
WTIMER=60.0s, IVOLT=0.5kV, IHIGH=OFF, ILOW=10MOHM, IMASK=1.0s,
ITIMER=60.0s, DISCHARGE=ON
○
When
FORMAT=OFF
SET:IW, 2.50, 1.50, 20.0, OFF, 60.0, 0.5, OFF, 10, 1.0, 60.0,ON
When the test mode is W,
○
When
FORMAT=ON
SET:MODE=W, WMARK=2.50kV, WLEVEL=1.50kV, WHIGH=20.0mA, WLOW=OFF,
WTIMER=60.0s
○
When
FORMAT=OFF
SET:W, 2.50, 1.50, 20.0, OFF, 60.0