17
Explanation of command
4.2.17
WLEVEL=
(setting of referential voltage of withstanding voltage test)
Function
Makes the setting of referential voltage of withstanding voltage test.
Structure
WLEVEL=
Referential voltage
Referential voltage OFF or 0.00~5.00kV is to be set.
Transmission
WLEVEL=1.50kV
…..
Sets the referential voltage of withstanding voltage test at
1.50kV.
Response
When 8527 received the effective command setting.
ERROR=0
…….……… When the Response Setting is ON.
No response ………………. When the Response Setting is OFF.
4.2.18
WLEVEL?
(read-out of referential voltage of withstanding voltage test)
Function
Reads out the referential voltage of withstanding voltage test.
Structure
WLEVEL?
Transmission
WLEVEL?
Response
WLEVEL=1.50kV
……. Indicates the referential voltage of withstanding voltage test
1.50kV.