21
Explanation of command
4.2.25
IVOLT=
(setting of test voltage range of insulation resistance test)
Function
Makes the setting of test voltage range of insulation resistance test.
Structure
IVOLT=
Test voltage range
Test voltage range 0.5kV or 1.0kV is to be set.
Transmission
IVOLT=1.0kV
………. Sets the range of insulation resistance test at 1.0kV.
Response
When 8527 received the effective command setting.
ERROR=0
…….……… When the Response Setting is ON.
No response ………………. When the Response Setting is OFF.
4.2.26
IVOLT?
(read-out of test voltage range of insulation resistance test)
Function
Reads out the test voltage range of insulation resistance test.
Structure
IVOLT?
Transmission
IVOLT?
Response
IVOLT=0.5kV
………. Indicates the test voltage range of insulation resistance test
0.5kV.