50
Remote control
14.4
●
Operation by REAR:MODE
Features of REAR:MODE
1. The test mode (withstanding voltage or insulation resistance test) can be selected by a
relay, sequencer etc. When the test mode is not selected, the test is performed by the
test condition before entering the REAR:MODE.
2. The test can be done, reading out the content of memory setting by a sequencer etc.
3. The test mode can be externally controlled but the change of numeral setting is not
possible, so make the setting in advance by the memory etc.
4. Since the tester is used by the external control, the tester becomes key lock condition
during the setting.
5. The start signal is decided depending upon the setting condition of remote control.
6. An interruption of the test is possible from the
STOP
switch
②
,
STOP
terminal
, on the rear panel and
Pin No.4 (STOP) of the
REMOTE/OUT
connector.
REAR:MODE from the setting to the start
(1) Make a short-circuit between the Pin No.20 (REAR:MODE) of the
REMOTE/OUT
connector
on the rear and COM (either 19, 23 or 36) of the same connector
.
is displayed on the memory number display.
Note
: When auto operation is done by the sequencer etc. without using the
START
switch
③
(manual start), make the Pin No.2 (REAR:ON) ON. Or, make a
short-circuit between REAR:ON and C of the
REMOTE
terminal
.
(2) Select a test mode.
Make a selection of either test mode in advance, withstanding voltage or insulation
resistance test, by means of Pin No.21 (W-MODE) or Pin No.22 (I-MODE) on the
REMOTE/OUT
connector
.
(3) After confirming the wiring with the test sample, safety and so on, press
START
switch
③
. Or, start the test by remote control.
In order to do the withstanding voltage and insulation resistance test in sequence (W-I,
I-W), make a re-start switching over the other test mode, after the good judgement at the
item (2) above.
[Example] When the withstanding voltage test
→
insulation resistance test (W-I) is done.
①
In READY status, turn ON the Pin No.21 (W-MODE) of the
REMOTE/OUT
connector
. The test is started. The tester becomes in operation of
withstanding voltage test and the
DANGER
lamp
⑬
is lit up.
②
After the good judgement of withstanding voltage test, turn OFF the Pin No.21
(W-MODE). The tester is then in READY status.
③
To do the insulation resistance test next, turn ON the Pin No.22 (I-MODE).
④
Start the test. The tester becomes in operation of insulation resistance test and
the
DANGER
lamp
⑬
is lit up.
⑤
Afterwards, the judgement can be made normally.
To start reading out the memory
(1) Make a short-circuit between the Pin No.20 (REAR:MODE) of the
REMOTE/OUT
connector
on the rear and COM (either 19, 23 or 36) of the same connector
.
is displayed on the memory number display.
(2) By the combination of the BCD code of the Pin No.6~9 (MEM SET 1, 2, 4, 8) of the
same connector
, read out the memory No.1~9.
Note
: When the A~F code is input, A~F is displayed on the display but no read out
is possible.
(3) After confirming the wiring with the test sample, safety and so on, press
START
switch
③
. Or, start the test by remote control.
Remote control which can be jointly used with REAR:MODE
Basically, it is as explained at the REAR:MODE from the setting to the start .
During the REAR:MODE setting, the remote control can also be used jointly.
The start from the
REMOTE
connector
⑧
(front panel),
REMOTE
terminal
(rear
panel) and Pin No.3 (START) of the
REMOTE/OUT
connector
is also possible.
Refer to the article 14.6 for the priority of remote control.
[Likely error at the REAR:MODE]
Blinking display of
For a likely cause and solution, refer to the
article 20 Error messages.
Blinking display of
Blinking display of