19
Explanation of command
4.2.21
WLOW=
(setting of low limit of leak current of withstanding voltage test)
Function
Makes the setting of low limit of leak current of withstanding voltage
test.
Structure
WLOW=
Low leak current
Low leak current OFF or 0.0~109.0mA is to be set.
Note
: Set value of low leak current can not be higher than high limit
value of leak current.
Transmission
WLOW=2.0mA
………... Sets the low limit of leak current of withstanding voltage
test at 2.0mA.
Response
When 8527 received the effective command setting.
ERROR=0
…….……… When the Response Setting is ON.
No response ………………. When the Response Setting is OFF.
4.2.22
WLOW?
(read-out of low limit value of leak current of withstanding voltage test)
Function
Reads out the low limit value of leak current of withstanding voltage
test.
Structure
WLOW?
Transmission
WLOW?
Response
WLOW=2.0mA
………... Indicates the low limit of leak current of withstanding
voltage test 2.0mA.