![Tsuruga 8527 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/tsuruga/8527/8527_instruction-manual_1178555038.webp)
38
10. Memory function
This tester is provided with 9 program memories to memorize the test mode and the
setting of test condition of withstanding voltage and insulation resistance test.
10.1
●
Configuration of memory
Each memory can memorize one type of test mode, 5 items of the test conditions of
withstanding voltage test and 6 items of the test conditions of insulation resistance test.
For the content of the type and items, please refer to the following table.
Type of test
mode
Setting of test condition for
withstanding voltage test
Setting of test condition for
Insulation resistance test
↓
↓
↓
1 items
5 items
6 items
W
→
I
I
→
W
W
I
Slider voltage setting
Range of test voltage
Referential voltage
High limit of resistance value
High limit of leak current
Low limit of resistance value
Low limit of leak current
Mask timer time
Test time
Test time
Discharging function
10.2
●
Memory write-in
Selection of memory No.
①
Make the setting of test mode and test condition required to be written in the memory,
and make the tester READY status (refer to the article 8~9).
②
Press
WRITE
key (
SHIFT
and
at a time), then the numeral on the memory
No. display blinks, entering into the memory write-in mode (refer to the above figure).
③
Select the memory No. to write in with
▲
or
▼
key.
Finish of memory write-in
Press
ENTER
key, then the tester returns to READY status, memorizing the set having
been made.
The slider moves up or down when the test mode is W-I or I-W (refer to the article 10.4).
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the memory write-in mode is interrupted and the tester becomes READY status.
The memory No. in this case is that before entering the memory write-in mode.
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
M
Ω
LOW
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
○
○
○
I
I-TEST
I
I-W
W-I
●
●
mA
HIGH
VOLTAGE SLIDER