22
8. Setting of test condition for withstanding voltage test
The test condition can be set when the test mode W-I, I-W or W is selected.
8.1
●
Voltage of motor slider
Adjustable range: 0.00~5.99kV, OFF
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks.
Test mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
W-I
,
I-W
or
W
lamp).
②
Press
key and make the test mode lamp lit.
Setting of slider voltage
①
Press
or
key and select the status that the
W AC
lamp is lit up, the test
voltage range display is lit up with
and the test voltage display blinks (refer to
the above figure).
②
While the test voltage display blinks, press
▲
or
▼
key and set the slider
voltage.
The second digit (0.10kV digit) can be set by pressing
key (
SHIFT
and
▲
at a time) or
key (
SHIFT
and
▼
at a time).
③
When the slider voltage setting is unnecessary, press
ON/OFF
key and select the
blinking
on the display.
To move to the previous setting
Press
key and the setting changes as follows depending upon the test mode.
Test mode
Setting item after change
W-I
→
Returns to the blinking of
W-I
test mode lamp.
W
→
Returns to the blinking of
W
test mode lamp.
I-W
→
To discharge of insulation resistance.
To the next setting
Press
key and it changes to the
setting of referential voltage
.
Finish of setting
Press
ENTER
key, then the slider moves up or down and the tester returns to READY
status, memorizing the set having been made.
When the
EXIT
key (
SHIFT
and
ENTER
at a time) is pressed in the setting mode,
the setting mode for the test condition is interrupted and the tester becomes READY
status.
The test conditions in this case are those before entering the setting mode of test condition.
ON/OFF
ENTER
WRITE
READ
LOCK
○
○
BUZZER
W
READY
REMOTE
DISCHARGE
KEY LOCK
EXIT
AC TEST VOLTAGE
MODE
JUDGE
TIME
RANGE
CURRENT/RESISTANCE
SHIFT
M
Ω
LOW
GOOD
HIGH
LOW
GOOD
HIGH
LOW
○
W-TEST
○
○
○
○
○
DC
AC
W
M
MEMORY
SET
SET
kV
s
kV
No.
○
○
○
I
I-TEST
I
I-W
W-I
mA
HIGH
●
●
VOLTAGE SLIDER