Preparation prior to use
Slider action at powering on after the setting
[at ON]
●
The display on the front panel becomes lamp test status.
●
The slider display goes down to 0.00kV. The tester enters the lamp test status. The
lamp test for the
DANGER
lamp is for about 3 seconds.
Caution
:
For the control signal and command during the process from powering on to the slider
voltage 0.00kV.
①
The
START
,
STOP
switches or keys are not accepted.
②
It becomes 0.00kV no matter whether the
REMOTE/OUT
connector is or not, or
regardless of set conditions.
③
The command of RS-232C is not accepted either.
[at OFF]
●
When the slider voltage is set in the test conditions before the power shutdown, the
slider goes up when the tester is powered ON and after finishing the lamp test.