![Tsuruga 8527 Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/tsuruga/8527/8527_instruction-manual_1178555076.webp)
8
4. Explanation of command
4.1
●
Table of command
Function
Setting / read-out
Approx.
response
time (ms)
(Note-1)
Expla-
nation
page
ON/OFF selection of remote control
REMOTE=/REMOTE?
23/19 9
Keylock
KEYLOCK=/KEYLOCK?
27/23 10
ON/OFF selection to suffix command name and
unit to the transmission to the host
FORMAT=/FORMAT?
27/23 11
ON/OFF selection of response
RESPONSE=/RESPONSE?
32/24 12
Test mode
MODE=/MODE?
17/15 13
Start of test
START
10~15 14
Stop of test and judgement reset
RESET
10~15 14
Read-out of status
STATUS?
5~13 15
Read-out of tester identification
IDNT?
12 16
Slider voltage
Note-2
WMARK=/ WMARK?
400/20 16
Referential voltage
WLEVEL=/WLEVEL?
28/16 17
High limit of leak current
WHIGH=/WHIGH?
25/16 18
Low limit of leak current
WLOW=/WLOW?
32/15 19
W
ithsta
nd-
ing vol
ta
ge
te
st
Test time
WTIMER=/WTIMER?
29/26 20
Test voltage range
IVOLT=/IVOLT?
19/16 21
High limit of resistance value
IHIGH=/IHIGH?
32/14 22
Low limit of resistance value
ILOW=/ILOW?
29/12 23
Mask time
IMASK=/IMASK?
30/18 24
Test time
ITIMER=/ITIMER?
25/18 25
In
sula
tion
res
is
tan
ce
te
st
Discharge function
DISCHARGE=/DISCHARGE?
32/27 26
Read-out of judgement result
JUDGE?
20 27
Read-out in the lump of test result and data
DATA?
16 28
Parameter of test condition
Note-2
SET:/SET:?
400/30 29
Change-over of memory No.
Note-2
MEMORY=/MEMORY?
400/14 30
Parameter of test condition including
memory No.
□
:1~9
MEM□:/MEM□:?
420/20 31
Buzzer sound volume
BUZZ=/BUZZ?
23/15 32
Note-1
: The response time mentioned in the table is the referential value and may vary
depending upon the condition of use. It is not to warrant the performance of
8527.
Note-2
: The setting command automatically moves the slider up or down.
Example :
WMARK=1.50kV
MEMORY=1
(When the effective
WMARK=□.□□kV
are in the memory
content.)
Host
C
8527
R
Response time