![Tsuruga 8527 Скачать руководство пользователя страница 93](http://html.mh-extra.com/html/tsuruga/8527/8527_instruction-manual_1178555093.webp)
25
Explanation of command
4.2.33
ITIMER=
(setting of test time of insulation resistance test)
Function
Makes the setting of test time of insulation resistance test.
Structure
ITIMER=
Test time
Test time OFF or 0.5~999 sec. is to be set.
Note
: Set the test time longer than mask time (IMASK) by 0.2 sec. or
more.
Transmission
ITIMER=60.0s
……... Sets the test time of insulation resistance test at 60.0 sec..
Response
When 8527 received the effective command setting.
ERROR=0
…….……… When the Response Setting is ON.
No response ………………. When the Response Setting is OFF.
4.2.34
ITIMER?
(read-out of test time of insulation resistance test)
Function
Reads out the test time of insulation resistance test.
Structure
ITIMER?
Transmission
ITIMER?
Response
ITIMER=10.0s
……... Indicates the test time of insulation resistance test 10.0 sec..