Information on ordering additional equipment for the operating table
88
4990113_030_01 — 2069106 — 2017–11–02
22
Information on ordering additional equipment for the operating table
Trumpf Medical offers a wide variety of section segments and
accessories for additional equipment of the operating table. For
more information, contact a Trumpf Medical representative (see
back of this document). Products used in conjunction with the
operating table (e.g. leg plates, radial control pistons) must be used
in compliance with their respective user manuals.
23
Radio license
Information on radio license is included in the instruction manual for
the radio module (document 4990136).
Table 6 according to IEC 60601-1-2:2007
Recommended separation distances between portable and mobile HF telecommunication devices and the
TruSystem 7500 operating table
The TruSystem 7500 operating table is intended for use in an electromagnetic environment where HF disturbance variables
are controlled. The customer or user of the above mentioned device can contribute towards avoiding electromagnetic
interference by observing the minimum distance between portable and mobile HF telecommunication devices (transmitters)
and the above mentioned device depending on the output power of the communication device as stated below.
Rated power of
transmitter
W
Safe distance depending on the transmitter frequency
150 kHz to 80 MHz
d=1.17
√
P
80 MHz to 800 MHz
d=1.17
√
P
800 MHz to 2.5 GHz
d=2.33
√
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters with a maximum nominal power not found in the table above, the recommended safe distance d can be
calculated in meters (m) using the equation for the respective column, where P is the maximum nominal power of the transmitter
in watts (W) according to the transmitter manufacturer's data.
Note 1
For 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines might not be applicable in all cases. The propagation of electromagnetic
quantities is influenced by the absorptions and reflections of buildings, items and human beings.