Power supply
42
4990113_030_01 — 2069106 — 2017–11–02
5
Power supply
5.1
Internal power supply (not available for TruSystem 7500 Hybrid MR)
5.1.1
mobile operating table column
Two lithium ion batteries provide the internal power supply for the
mobile operating table column. With the set of batteries fully
charged, the use of the electric operating table functions is
guaranteed for approx. 1 hour (total duration of movement for
electrical functions). The electrical functions on the operating table
are blocked when the rechargeable battery set is depleted or
almost completely depleted. In this case, the external power supply
must be connected.
The battery charge status can be seen on the display [a1] on the
column keypad. Observe the display while using the operating
table.
5.1.2
Fixed installation operating table column
Two lithium ion batteries provide the internal power supply for the
fixed installation operating table column. With the set of batteries
fully charged, the use of the electric operating table functions is
guaranteed for approx. 30 minutes (total duration of movement for
electrical functions). The battery is only used as an emergency
power supply in the event of a power failure. During battery
operation, the electrical functions on the operating table are moved
at a lower speed. The electrical functions on the operating table are
blocked when the rechargeable battery set is depleted or almost
completely depleted. Due to the short operating time of the battery,
the external power supply must be restored as quickly as possible.
The battery charge status can be seen on the display [a1] on the
column keypad. Observe the display while using the operating
table.