Configuration options
4990113_030_01 — 2069106 — 2017–11–02
63
13
Configuration options
The following settings can be configured via the service interface of
the operating table:
1. Stop on passing through the level position (activated for all
functions by default):
All electrical adjustment functions of the operating table stop
automatically when the level position is reached. For any
further adjustment, briefly release the function key and then
press it again.
2. Restricted leveling (not active by default):
With the level position function, the operating tabletop
motorized hinges set upward remain in position and are no
longer leveled.
3. Interim stop for table tilt, Trendelenburg and longitudinal travel
(not activated by default):
An intermediate stop can be established for the tilting,
Trendelenburg, and longitudinal slide functions. The function
stops automatically upon reaching the position set individually,
and not at the level position. A message appears on the
TruSystem 7500 remote control display. For additional
adjustment beyond the threshold value, briefly release the
function key and press it again. For safety reasons, the tilting
intermediate stop cannot be deactivated (intermediate stop at
a tilt angle of <15° is possible).
4. Sensor-supported transfer (activated by default):
Extreme centers of gravity outside the center can prevent the
complete insertion of the mount wedges (column) into the
wedge receptacles of the operating tabletop. With the help of
the sensor-supported transfer, the center of gravity of the
operating tabletop is automatically optimized during the
transfer, so that the mount wedges are easier to engage.
5. Signal tone volume:
Signal tone volume can be set individually at the operating
table.
The configuration is carried out via the service interface of the
operating table (Ethernet connection from the operating table to the
PC). The desired option has to be set on-site by the Trumpf Medical
Technical Customer Service or the responsible hospital technician.