Troubleshooting
80
4990113_030_01 — 2069106 — 2017–11–02
Back section cannot be adjusted
far enough down
Longitudinal travel too far toward
the foot end
Adjust longitudinal travel toward
the head end
Acoustic signal during leveling,
when leg sections move together
Possible risk of collision (signal is
warning)
OK (not an error)
Transfer of the operating tabletop
from the operating table column
to a shuttle is not possible
Drive defective
Prepare the emergency transfer
(see chapter14.4 on page71).
Failure of all operating table
functions
Central operating table control is
defective
Activate emergency mode (see
chapter12.7 on page59).
Error
Possible cause
Correction