Control units
36
4990113_030_01 — 2069106 — 2017–11–02
4.2
Column keypad
4.2.1
Overview of functions
The operating table is shown in simplified form on the column
keypad: in the upper section, the operating tabletop (longitudinal
side) and underneath, the operating table column as viewed from
the head end. The keys for the adjustment functions are identified by
arrows and located on the corresponding table segments. The
direction of movement for the individual functions corresponds to the
image displayed.
The individual functions on the column keyboard can only be
selected once the keypad is unlocked. To adjust an operating table,
press and hold the function key on the keypad until the desired
position is reached. The function stops when the key is released. The
keypad is locked again automatically ten seconds after the last key
press. The
O
FF
button is an exception. The operating table can be
turned off at any time using the
O
FF
key on the column keypad.
The column keypad functions can also be selected using two-key
control. Press the
O
N
key and simultaneously press the desired
function key. The keypad does not need to be unlocked prior to
two-key control.
Key:
[i1] Switch on the operating table, display: Ready
[i2] Switch off the operating table
[i3] Head position left, display: Head position left
[i4] Head position right, display: Head position right
[i5] Lift
[i6] Lower
[i7] Trendelenburg
[i8] Reverse Trendelenburg
[i9] Tilt left
[i10] Tilt right
Содержание 1717020