5.6.1.2.1. Control Relay 1 and 2 Temperature function menu
Line 1
T E M P E R A T U R E F U N C T I O N
Line 2
T e m p . - O k :
- 8 8 K
Line 3
5 - 9 9 K + 8 8 K
Line 4
S t a b . t i m e : 8 8 . 8 s
Line 5
0 - 9 9 . 9 s
Line 6
H o l d :
O f f
Line 7
O n
Line 8
Press the button
to return to the Control Relay settings
menu (
5.6.1.2.). Here, the control relay 1 and 2 with the
function “Temperature Ok-message” can be set.
Line 1:
Menu name
Line 2…3:
Setting of the upper and lower limit of the tem-
perature Ok-range around the desired value. The adjustable
range is shown in line 3.
Line 4…5:
Setting of the stabilisation time of the tempera-
ture Ok message. The adjustable range is shown in line 5.
Line 6…7:
Options for switching the holding function of the
control relay on and off. The relay remains actuated after a
welding cycle up to the start of the following welding cycle.
5.6.1.2.2. Control Relay 1 and 2 Sequence function menu
Line 1
S E Q U E N C E F U N C T I O N
Line 2
A c t . w i t h s t a r t s i g .
Line 3
A c t . w i t h t e m p . r e a .
Line 4
A c t . d u r . h e a t . p h a s e
Line 5
A c t . d u r . c o o l . p h a s e
Line 6
C a l i b r a t i o n O k
Line 7
A l a r m
Line 8
T c c o r r . s e q u e n c e
Press the button
to return to the Control Relay settings
menu (
5.6.1.2.). Here, the control relay 1 and 2 can be
set.
Line 1:
Menu name
Line 2:
Option that the control relay picks up when the
“Start” signal is applied and is released when the cooling
phase has elapsed.
Line 3:
Option that the control relay picks up when the ac-
tual value has reached 95 % of the desired value and is
released when the cooling phase has elapsed.
Line 4:
Option that the control relay is actuated during the
heating phase.
Line 5:
Option that the control relay is actuated during the
cooling phase.
Line 6:
Option that the control relay is actuated after a suc-
cessful calibration.
Line 7:
Option that the control relay is actuated in the error
state (from V1.00/1.19/1.16/1.10).
Line 8:
Option that the control relay performs the control
message for the temperature coefficient correction (from
V1.00/1.23/1.30/1.25)
5.6.1.2.3. Control Relay 1 and 2 Time Pulse function menu
Line 1
T I M E P U L S E F U N C T I O N
Line 2
P r e h e a t . p h . : 8 8 . 8 s
Line 3
0 - 9 9 . 9 s
Line 4
C o o l . p h a s e : 8 8 . 8 s
Line 5
0 - 9 9 . 9 s
Line 6
E n d o f p u l s e
Line 7
Line 8
Press the button
to return to the Control Relay settings
menu (
5.6.1.2.). Here, the control relay 1 and 2 can be
set.
Line 1:
Menu name
Line 2…3:
Option that the control relay is actuated for the
preset time when the welding process or the welding time is
started. The adjustable range is shown in line 3.
Line 4…5:
Option that the control relay is actuated for the
preset time when the cooling phase (end of the heating
phase) is started. The adjustable range is shown in line 5.
Line 6:
Option that the control relay is actuated for 500ms
at the end of the cooling phase.
5.6.1.3. Alarm Relay settings menu
Line 1
A L A R M R E L A Y
Line 2
O o n t a c t :
N O C
Line 3
N C
Line 4
Line 5
Line 6
Line 7
Line 8
Press the button
to return to the Relay settings menu (
5.6.1.).
Line 1:
Menu name
Line 2…3:
Options for the switching mode of the relay, in
case of an error either as NO or NC.