3.7.1. Possible errors:
Table 1
No. Error
Actual
value
output
Corrective action
10 Device error
4.66 / 0 V
perform reset
21 internal error
4.00 V
perform reset
22 read-write error in the non-volat. Memory
31 mains voltage too low
3.33 V
check mains connection and
perform reset
32 mains voltage too high
33 mains frequency error
40 current signal Ir and voltage signal Ur too
low
2.00 V
perform calibration and
check heating circuit
50 voltage signal Ur too low
1.33 V
perform connection to voltage measure-
ment Ur and calibration
60 current signal Ir too low
0.66 V
check connection to current measurement
Ir and perform calibration
71 current signal Ir too high
5.33<>10 V
check heat conductor and
perform calibration
72 voltage signal Ur too high
check connection to voltage measure-
ment Ur and perform calibration
73 current sig. Ir and voltage sig. Ur too high
perform calibration
81 heat conductor temperature too high
2.66 V
check heat conductor and
perform calibration
82 heat conductor temperature too low
83 temperature monitor (
3.6.1.) tempera-
ture too high
check setting,
check heat conductor and
perform calibration
84 temperature monitor (
3.6.1.) tempera-
ture too low
85 heating monitor (
3.6.2.)
87 temperature step down
(from V1.00/1.07/1.29/1.23)
loose contact,
check heating circuit
88 temperature step up
(from V1.00/1.07/1.29/1.23)
90 data error - saved calibration values do
not match setting
6.00<>10 V
perform calibration
91 communication monitor RS232 interface
(
3.6.4)
check RS232 interface connection
92 internal communication monitor 1
perform reset
93 communication monitor USB interface (
3.6.4)
check USB interface connection
94 internal communication monitor 2
perform reset
95 communication monitor RS232 interface
for exTM thermometer
check RS232 interface connection
and exTM thermometer
Calibration error:
101 current signal Ir and voltage signal Ur too
high
8.00<>10 V
check dimensioning
102 current signal Ir and voltage signal Ur too
low
check heat conductor connection and
dimensioning
103 current signal Ir too low and voltage signal
Ur too high
check dimensioning
104 current signal Ir too high and voltage sig-
nal Ur too low
105 R20 cannot be detected
106 P-factor unable to be determined
111 voltage signal Ur too low
7.33<>10 V
check connection to voltage measure-
ment Ur and dimensioning
112 voltage signal Ur too high
check connection to voltage measure-
ment Ur, heat conductor and
dimensioning
113 voltage signal Ur unstable
(from V1.00/1.01/1.19/1.18)
check connection to voltage measure-
ment Ur