6 SPECIAL MODE (Menu Mode)
6.4.3 Vertical and Horizontal Angle Tilt correction (Tilt ON/OFF)
In case the instrument is used in an unstable situation, constant indexing of vertical and horizontal
angle may be impossible. In this case, the function of tilt correction can be stopped by selecting TILT
OFF. It has been set to X,Y (V/H) TILT ON at the factory.
•
This setting is memorized after power off.
6.4.4 Systematic Error of Instrument Correction
Setting ON/OFF for error correction of collimation and horizontal axis for angle measurement.
•
Perform this item after complete Section 18.3 “Adjustment of Compensation Systematic Error of
Instrument”.
Operating procedure
Operation
Display
1
After pressing the
{MENU}
key, press the
{F4}
(P
↓
) key twice to get the menu on page 3.
{MENU}
{F4}
{F4}
2
Press the
{F2}
key.
{F2}
3
Press the
{F3}
key.
The data previously set is shown.
If already ON, the display shows tilt correction
value.
{F3}
4
Press the
{F1}
(X-ON) key or
{F2}
(XY-ON) or
{F3}
(OFF) key, and press the
{F4}
(ENTER) key.
{F1}
~
{F3}
{F4}
Operating procedure
Operation
Display
1
After pressing the
{MENU}
key, press the
{F4}
(P
↓
) key twice to get the menu on page 2.
{MENU}
{F4}
{F4}
2
Press the
{F2}
key.
{F2}
3
Press the
{F4}
key.
{F4}
MENU 3/4
F1:ILLUMINATION
F2:PARAMETERS 1
F3:PARAMETERS 2 P
↓
PARAMETERS 1 1/3
F1:MINIMUM READING
F2:AUTO POWER OFF
F3:TILT P
↓
TILT SENSOR:[XY-ON ]
X: 0°02'10"
Y: 0°03'00"
X-ON XY-ON OFF ENTER
MENU 3/4
F1:ILLUMINATION
F2:PARAMETERS 1
F3:PARAMETERS 2 P
↓
PARAMETERS 1 1/3
F1:MINIMUM READING
F2:AUTO POWER OFF
F3:TILT P
↓
PARAMETERS 1 2/3
F1:ERROR CORRECTION
F2:RS-232C
F3:SELECT COM P
↓