12-2
12 SETTING ATMOSPHERIC CORRECTION
12.2 Setting of Atmospheric Correction Value
How to Set Temperature and Pressure Value Directly
Measure the temperature and air pressure surrounding the instrument beforehand.
Example: Temperature: 26°C, Pressure:1017 hPa
Operating procedure
Operation
Display
1
Press the
{F4}
key to get the soft key page 2 in the
distance measurement mode or page 3 in the
coordinate measurement.
{F4}
2
Press the
{F3}
(S/A) key to set Set Audio Mode
from distance or coordinate measurement mode.
{F3}
3
Press the
{F3}
(T-P) key.
{F3}
4
Input Temperature value and Pressure value.*1)
Mode returns to Set Audio mode.
Enter Temp.
Enter Pres.
*1)
Refer to Section 2.4 “How to Enter Alphanumeric Characters”.
Range: Temp. –30 to 60°C (0.1°C step) or –22 to 140°F (0.1°F step)
Pres. 560 to 1066.0hPa (0.1hPa step), 420 to 800mmHg (0.1mmHg step) or
16.5 to 31.5inHg (0.1inHg step)
HR: 120°30'40"
HD* 123.456 m
VD: 5.678 m
MEAS MODE NP/P P1
↓
OFSET S.O S/A P2
↓
PSM:0.0 PPM 0.0
NPM:0.0 SHT: 0.0
SIGNAL:[
IIIII
]
PRISM PPM T-P –––
TEMP. & PRES. SET
TEMP. =15.0 °C
PRES. : 1013.2 hPa
--- --- [CLR] [ENT]
TEMP. & PRES. SET
TEMP. : 26.0 °C
PRES. : 1017.0 hPa
INPUT ––– ––– ENTER