24-2
24 ERROR DISPLAYS
FULL
FILES
When making a file, 30 files already exist.
If necessary, send or delete files.
FAILED
INITIALIZE
Initializing can not be done successful.
Confirm initializing data and try to initialize
again.
LIMIT
OVER
Limit of input data exceeded.
Input again.
MEMORY
ERROR
Any abnormality occurs with internal memory. Initialize the internal memory.
MEMORY
POOR
Shortage of capacity of the internal memory.
Download data from internal memory to PC.
NO DATA
The data is not found in the search mode
Confirm the data and search again.
NO FILE
There is no file in internal memory.
If necessary, make files.
FILE NOT
SELECTED
When using a file, no file is selected.
Confirm the file and select a file.
P1-P2
distance
too short
When in point to line measurement, the
horizontal distance between first point and
second point is within 1m.
The horizontal distance between first point
and second point must be more than 1m.
CIRCULAR
ERROR
Known points and occupied point are on the
same circle at the resection mode.
Take the different point.
PT# EXIST
Same new point name is already memorized
in the memory.
Confirm the new point name and input again.
PT# DOES
NOT EXIST
When you enter incorrect name or PT# does
not exist in the internal memory.
Enter the correct name or enter point in the
internal memory.
RANGE
ERROR
When setting a new point, calculation is
impossible from the measured data.
Measure again.
READ ONLY
FILE
Read-only file on the USB flash drive cannot be
changed and file contents cannot be deleted or
edited.
Remove the read -only attribute of the file on
your PC etc.
Temp
RangeOUT
ES is outside useable temperature range and
accurate measurement cannot be performed.
Repeat measurement within the appropriate
temperature range. If the ES is used under
direct sunlight, use an umbrella to protect it
against the heat of the sun.
Tilt Over
Instrument tilts over more than 3 minutes.
Level the instrument properly.
USB ERROR
An error has occurred in loading or saving
data to USB flash drive.
Check if the USB flash drive is formatted to
FAT32.
Insert the USB flash drive correctly.
If the error still occurs, use another USB flash
drive.
V ANGLE
ERROR
H ANGLE
ERROR
VH ANGLE
ERROR
Abnormality in angle measuring system.
If this error code continues to display, repair
is required.