24-1
24 ERROR DISPLAYS
24
ERROR DISPLAYS
The following is a list of the error messages displayed by the instrument and the meaning of each
message.
•
There is a possibility that an unexpected error occasionally occurs when using an old battery. In such a
case, exchange it for a new battery, and please check the operation of the instrument.
•
If error still persist after attempting to clear them, contact your local dealer.
Error code
Description
Countermeasures
3 points
required
When area calculating, there are less than 3
points coordinate data in selected file.
Confirm the file data and calculate again.
BadCondition
The air is shimmering a lot, etc., measuring
conditions are poor.
The center of the target cannot be sighted.
Unsuitable distance measurement conditions
when reflectorless measurement is set.
When reflectorless measurement is set,
distance cannot be measured because the
laser beam is striking at least two surfaces at
the same time.
Perform measurement again when
measuring condition becomes better.
Resight the target.
Choose a single surface target for distance
measurement.
CALC
ERROR
Calculation is impossible from the data input.
Confirm the input data.
DELETE
ERROR
When deleting coordinate data, it can not
be done successful.
Confirm the data and delete again.
SIGNAL
OFF
Measuring conditions are poor, and there is
not enough reflective light for measuring
distances.
Resight the target. When using reflective
prisms, effectiveness will be improved by
increasing the number of prisms used.
E020’s,
E031, E033
EDM error
Repair is required.
E35
Displayed when REM measurement carried
out to the range from zenith or nadir ±6°.
Operate in the range out of ± 6° from the
zenith or nadir.
E72
Displayed when Vertical angle -position is
adjusted in wrong position.
Repair is required.
E73
The instrument was not leveled when Vertical
angle 0-position is adjusted.
Level the instrument then carry the
adjustment work.
E80’s
Mainly at the time data transmission between
ES and external instrument.
Confirm operation procedure is correct or the
connection cables are correct.
E097
EDM device error
Repair is required.
E123, E124
Tilt error
Contact your local dealer.
E198
Angle measurement error
Contact your local dealer.
E201
EDM distance measurement time out error
If the same error message is repeated, repair
is required.
E205
Communication with EDM is poor
If the same error message is repeated, repair
is required.
FILE
EXISTS
The same file name exists.
Use another file name.