172-65710M-04 (SC-F71 Parameters/Functions) 23 Dec 2021
112
K
K02
0 to 400
C
4
T
T02
-199.9 to +100.0
C
4
K03
0 to 600
C
4
T03
-100.0 to +200.0
C
4
K04
0 to 800
C
4
T19
-200.0 to +400.0
C
5
K06
0 to 1200
C
4
R
R01
0 to 1600
C
4
K07
0 to 1372
C
4
R07
-50 to +1768
C
4
K08
-199.9 to +300.0
C
4
S
S06
-50 to +1768
C
4
K09
0.0 to 400.0
C
4
B
B03
0 to 1800
C
4
K10
0.0 to 800.0
C
4
E
E01
0 to 800
C
4
K14
0 to 300
C
4
N
N02
0 to 1300
C
4
K41
-200 to +1372
C
4
W5Re/
W26Re
W03
0 to 2300
C
4
K42
-200.0 to +1372.0
C
5
KA1
0 to 800
F
4
PL II
A01
0 to 1300
C
4
KA2
0 to 1600
F
4
U
U01
-199.9 to +600.0
C
4
KA3
0 to 2502
F
4
L
L04
0.0 to 900.0
C
4
J
J01
0 to 200
C
4
PR40-20
F02
0 to 1800
C
5
J02
0 to 400
C
4
FA2
0 to 3200
F
5
J03
0 to 600
C
4
J04
0 to 800
C
4
J08
0.0 to 400.0
C
4
J29
-200.0 to +1200.0
C
5
JA1
0 to 800
F
4
JA3
0 to 2192
F
4
JA6
0 to 400
F
4
RTD input
Input
type
Code
Range
No. of
digits**
Input
type
Code
Range
No. of
digits**
PT100
D01
-199.9 to +649.0
C
4
Pt100
D21
-200.0 to +200.0
C
5
D04
-100.0 to +100.0
C
4
D27
0.00 to +50.00
C
4
D05
-100.0 to +200.0
C
4
D34
-100.00 to +100.00
C
5
D06
0.0 to 50.0
C
4
D35
-200.0 to +850.0
C
5
D07
0.0 to 100.0
C
4
DA1
-199.9 to +999.9
F
4
D08
0.0 to 200.0
C
4
DA9
0.0 to 500.0
F
4
D09
0.0 to 300.0
C
4
JPt100
P08
0.0 to 200.0
C
4
D10
0.0 to 500.0
C
4
P29
-100.00 to +100.00
C
5
D12
-199.9 to +600.0
C
4
P30
-200.0 to +640.0
C
5
Voltage/Current input
Input type
Code
Range
No. of
digits**
Input type
Code
Range
No. of
digits
0 to 10 mV DC
101
Programmable
range
-199999 to +99999
5
1 to +5 V DC
601
Programmable
range
-19999 to +99999
5
0 to 100 mV DC
201
5
0 to 20 mA DC
701
5
0 to 1 V DC
301
5
4 to 20 mA DC
801
5
0 to 5 V DC
401
5
-10 to +10V DC
904
5
0 to 10 V DC
501
5
-5 to +5 V DC
905
5
Содержание SC-F71
Страница 16: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 15 2 loop control Cascade control Control with PV select ...
Страница 17: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 16 ...
Страница 18: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 17 Differential temperature control ...
Страница 114: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 113 Setting procedure ...
Страница 131: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 130 Setting procedure ...
Страница 134: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 133 Setting procedure ...
Страница 141: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 140 Setting procedure ...
Страница 153: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 152 Setting procedure ...
Страница 158: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 157 Setting procedure ...
Страница 168: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 167 Setting procedure ...
Страница 172: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 171 Setting procedure ...
Страница 175: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 174 Setting procedure ...
Страница 180: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 179 Setting procedure ...
Страница 185: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 184 ...
Страница 199: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 198 Setting procedure ...
Страница 207: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 206 Setting procedure ...
Страница 236: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 235 Setting procedure Selecting Heating Cooling PID control ...
Страница 249: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 248 Setting procedure ...
Страница 280: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 279 Setting procedure ...
Страница 293: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 292 Checking PID values ...
Страница 303: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 302 Setting procedure ...
Страница 316: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 315 Setting procedure ...
Страница 320: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 319 Setting procedure ...
Страница 347: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 346 Operating procedure ...
Страница 359: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 358 Setting Procedure Setting Soft start function ...
Страница 388: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 387 Setting procedure ...