172-65710M-04 (SC-F71 Parameters/Functions) 23 Dec 2021
111
One decimal place
-50.0 to +1768.0
C
-58.0 to +3214.0
F
E
No decimal place
-200 to +1000
C
-328 to +1832
F
One decimal place
-200.0 to +1000.0
C
-328.0 to +1832.0
F
B
No decimal place
0 to 1800
C
0 to +3272
F
One decimal place
0.0 to 1800.0
C
0.0 to +3272.0
F
N
No decimal place
0 to 1300
C
0 to +2372
F
One decimal place
0.0 to 1300.0
C
0.0 to +2372.0
F
W5Re/W26Re No decimal place
0 to 2300
C
0 to +4200
F
PL II
No decimal place
0 to 1390
C
0 to +2534
F
One decimal place
0.0 to 1390.0
C
0.0 to +2534.0
F
U
No decimal place
-200 to +600
C
-328 to +1112
F
One decimal place
-200.0 to +600.0
C
-328.0 to +1112.0
F
L
No decimal place
0 to 900
C
0 to 1652
F
One decimal place
0.0 to 900.0
C
0.0 to 1652.0
F
PR40-20
No decimal place
0 to 1800
C
0 to 3200
F
* The least significant digit (LSD) may flicker when the display resolution is set to 0.1 °C (0.1 °F).
K, J, T, S, R, E, B, N, PL II, U, and L are settable to one decimal place. Other
thermocouple inputs are fixed with no decimal place.
RTD input
Input type
Decimal point position
C
F
Pt100
No decimal place
-200 to +850
C
-328 to +1562
F
-100 to +100
C
-148 to +212
F
0 to 50
32 to 122
F
One decimal place
-200.0 to +850.0
C
-328.0 to +1562.0
F
-100.0 to +100.0
C
-148.0 to +212.0
F
0.0 to 50.0
32.0 to 122.0
F
Two decimal place
-100.00 to +100.00
C
-148.00 to +212.00
F
0.00 to +50.00
C
32.00 to +122.00
F
JPt100
No decimal place
-200 to +649
C
-328 to +1184
F
-200 to +100
C
-148 to +212
F
0 to 50
32 to 122
F
One decimal place
-200.0 to +649.0
C
-328.0 to +1184.0
F
-100.0 to +100.0
C
-148.0 to +212.0
F
0.0 to 50.0
32.0 to 122.0
F
Two decimal place
-100.00 to +100.00
C
-148.00 to +212.00
F
0.00 to 50.00
C
-32.00 to +122.00
F
Current/voltage input
Input type
Decimal point position
C
Pt100
No decimal place
-19999 to +99999
One decimal place
-1999.9 to +9999.9
Two decimal place
-199.99 to +999.99
Three decimal place
-19.999 to +99.999
Four decimal place
-1.9999 to +9.9999
Input range code table
The Input range code table is a list of input range codes so that a user can specify
the input range at the time of ordering. Even if the input range has been specified
when ordered, the input range can be changed later within the measured range.
Thermocouple (TC) input
Input
type
Code
Range
No. of
digits**
Input
type
Code
Range
No. of
digits**
K01
0 to 200
C
4
T01
-199.9 to +400.0
C
4
Содержание SC-F71
Страница 16: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 15 2 loop control Cascade control Control with PV select ...
Страница 17: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 16 ...
Страница 18: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 17 Differential temperature control ...
Страница 114: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 113 Setting procedure ...
Страница 131: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 130 Setting procedure ...
Страница 134: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 133 Setting procedure ...
Страница 141: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 140 Setting procedure ...
Страница 153: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 152 Setting procedure ...
Страница 158: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 157 Setting procedure ...
Страница 168: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 167 Setting procedure ...
Страница 172: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 171 Setting procedure ...
Страница 175: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 174 Setting procedure ...
Страница 180: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 179 Setting procedure ...
Страница 185: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 184 ...
Страница 199: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 198 Setting procedure ...
Страница 207: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 206 Setting procedure ...
Страница 236: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 235 Setting procedure Selecting Heating Cooling PID control ...
Страница 249: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 248 Setting procedure ...
Страница 280: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 279 Setting procedure ...
Страница 293: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 292 Checking PID values ...
Страница 303: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 302 Setting procedure ...
Страница 316: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 315 Setting procedure ...
Страница 320: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 319 Setting procedure ...
Страница 347: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 346 Operating procedure ...
Страница 359: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 358 Setting Procedure Setting Soft start function ...
Страница 388: ...172 65710M 04 SC F71 Parameters Functions 23 Dec 2021 387 Setting procedure ...